Dräger Tube Organic Basic Nitrogen Compounds - PN: CH25903
Dräger Tube Organic Basic Nitrogen Compounds
PN: CH25903
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of organic alkaline nitrogen compounds and has the following standard measuring range: 1 mg/m³ threshold value Measurements: 10
Application Range:
- Standard Measuring Range: 1 mg/m3 corresponds to a discoloration of 1 to 2 mm in length.
- Number of Strokes n: 8
- Time for Measurement: approx. 1.5 min
- Standard Deviation: ± 50 %
- Color Change: Yellow → Orange Red
Ambient Operating Conditions:
- Temperature: 0 to 40 °C
- Absolute Humidity: < 50 mg H2O / L
Reaction Principle:
- NR3 + KBiI4 → orange red reaction product
Cross Sensitivity:
- Various organic basic nitrogen compounds are indicated.
- It is impossible to differentiate between them.
Dräger Tube Organic Basic Nitrogen Compounds
PN: CH25903
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of organic alkaline nitrogen compounds and has the following standard measuring range: 1 mg/m³ threshold value Measurements: 10
Application Range:
- Standard Measuring Range: 1 mg/m3 corresponds to a discoloration of 1 to 2 mm in length.
- Number of Strokes n: 8
- Time for Measurement: approx. 1.5 min
- Standard Deviation: ± 50 %
- Color Change: Yellow → Orange Red
Ambient Operating Conditions:
- Temperature: 0 to 40 °C
- Absolute Humidity: < 50 mg H2O / L
Reaction Principle:
- NR3 + KBiI4 → orange red reaction product
Cross Sensitivity:
- Various organic basic nitrogen compounds are indicated.
- It is impossible to differentiate between them.