Dräger MT Xylene 10-1000 ppm - PN 8610260
Dräger MT Xylene 10-1000 ppm
PN : 8610260
|
Measuring Range |
10 to 1000 ppm |
| Typical Measuring Time |
10 to 240 s |
| Accuracy | ± 25% ( under calibration conditions ) |
|
Ambient Conditions |
|
| Temperature |
0 to 40 °C |
| Humidity |
1 to 40 mg/L |
|
Cross Sensitivity |
No influence at 50 ppm xylene by: 5 ppm toluene 5 ppm benzene 300 ppm n-octane |
| Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
| Hose |
Length ( meter ) |
Flashing time (min) |
| FKM hose |
5 - 20 |
3 |
| FKM hose |
45 |
20 |
| PTFE-line PVC hose |
3 - 15 | 3 |
| Bar probe 90 (83 16 532) |
5 | |
| Telescopic probe ES 150 ( 83 16 533) |
7 | |
| Probe set GP600 )83 28 667) | 3 | |
Dräger MT Xylene 10-1000 ppm
PN : 8610260
|
Measuring Range |
10 to 1000 ppm |
| Typical Measuring Time |
10 to 240 s |
| Accuracy | ± 25% ( under calibration conditions ) |
|
Ambient Conditions |
|
| Temperature |
0 to 40 °C |
| Humidity |
1 to 40 mg/L |
|
Cross Sensitivity |
No influence at 50 ppm xylene by: 5 ppm toluene 5 ppm benzene 300 ppm n-octane |
| Hose and probe measurement |
The following flushing time recommendations for hose and probe measurement are marked with a brand-new , dry and clean hose and probes. |
| Hose |
Length ( meter ) |
Flashing time (min) |
| FKM hose |
5 - 20 |
3 |
| FKM hose |
45 |
20 |
| PTFE-line PVC hose |
3 - 15 | 3 |
| Bar probe 90 (83 16 532) |
5 | |
| Telescopic probe ES 150 ( 83 16 533) |
7 | |
| Probe set GP600 )83 28 667) | 3 | |