Dräger-Tube Nitrous Fumes 0,2/a - PN: 8103661

$109.00
Image

Dräger-Tube Nitrous Fumes 0.2/a (10 Tests per Box)
PN: 8103661

Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of nitrous gases and has the following standard measuring ranges: 0.2-6 ppm Measurements: 10



Application Range

The first line on the tube scale corresponds to 0.2 ppm
Standard Measuring Range 2 to 60 ppm 
Number of Strokes n 5
Time for Measurement approx. 75 s
Standard Deviation ± 10 to 15%
Color Change gay green to blue gray
Ambient Operating Conditions
Temperature
10 to 40 °C
Absolute Humidity 3 to 4 mg H2O/L 
 Cross Sensitivity
In case of nitrogen dioxide in concentrations above 300 ppm  the indication layer can bleach. Chlorine and ozone are indicated as well, but with different sensitivities and may falsify the measurement result.

Dräger-Tube Nitrous Fumes 0.2/a (10 Tests per Box)
PN: 8103661

Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of nitrous gases and has the following standard measuring ranges: 0.2-6 ppm Measurements: 10



Application Range

The first line on the tube scale corresponds to 0.2 ppm
Standard Measuring Range 2 to 60 ppm 
Number of Strokes n 5
Time for Measurement approx. 75 s
Standard Deviation ± 10 to 15%
Color Change gay green to blue gray
Ambient Operating Conditions
Temperature
10 to 40 °C
Absolute Humidity 3 to 4 mg H2O/L 
 Cross Sensitivity
In case of nitrogen dioxide in concentrations above 300 ppm  the indication layer can bleach. Chlorine and ozone are indicated as well, but with different sensitivities and may falsify the measurement result.