Dräger Tube Natural Gas Test - PN: CH20001
Dräger Tube Natural Gas Test (5)
PN: CH20001
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of natural gas and has the following standard measuring range: 100-1000 ppm Measurements: 5
Application Range
Standard Measuring Range: Qualitative determination of natural gas
Number of Strokes n: 2
Time for Measurement: approx. 40 s
Color Change: white ä brown green to grey violet
Ambient Operating Conditions
Temperature: 0 to 50 °C
Absolute Humidity: max. 40 mg H2O / L
Reaction Principle
a) CH4 + KMnO4 + H2S2O7 ä CO
b) CO + I2O5 ä I 2 + CO2
Cross Sensitivity
Due to the reaction principle a number of other organic compounds such as propane and butane are indicated as well. Carbon monoxide is also indicated. It is not possible to differentiate between different compounds.
Dräger Tube Natural Gas Test (5)
PN: CH20001
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of natural gas and has the following standard measuring range: 100-1000 ppm Measurements: 5
Application Range
Standard Measuring Range: Qualitative determination of natural gas
Number of Strokes n: 2
Time for Measurement: approx. 40 s
Color Change: white ä brown green to grey violet
Ambient Operating Conditions
Temperature: 0 to 50 °C
Absolute Humidity: max. 40 mg H2O / L
Reaction Principle
a) CH4 + KMnO4 + H2S2O7 ä CO
b) CO + I2O5 ä I 2 + CO2
Cross Sensitivity
Due to the reaction principle a number of other organic compounds such as propane and butane are indicated as well. Carbon monoxide is also indicated. It is not possible to differentiate between different compounds.