Dräger Tube Methylene Chloride 20/a - PN: 8103591
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Methylene Chloride 20/a - PN: 8103591

$105.00
Image

Dräger Tube Methylene Chloride 20/a (10)
PN: 8103591

Application Range
Standard Measuring Range: 20 to 200 ppm
Number of Strokes n: 8
Time for Measurement: 7
Standard Deviation: ± 15 to 25 %
Color Change: yellow ä red

Ambient Operating Conditions
Temperature: 17 to 30 °C At 25 °C to 30 °C/77 °F to 86 °F multiply the reading with the factor 0.6.
Absolute Humidity: 3 to 25 mg H2O / L

Reaction Principle
a) CH2CI2 + chromateä CI2
b) CI2 + amine ä red reaction product

Cross Sensitivity
100 ppm n-octane and 300 ppm carbon monoxide do not interfere the reading. In case of concentrations > 100 ppm n-octane, methylene chloride is not indicated. Other chlorinated hydrocarbons will be indicated.

Dräger Tube Methylene Chloride 20/a (10)
PN: 8103591

Application Range
Standard Measuring Range: 20 to 200 ppm
Number of Strokes n: 8
Time for Measurement: 7
Standard Deviation: ± 15 to 25 %
Color Change: yellow ä red

Ambient Operating Conditions
Temperature: 17 to 30 °C At 25 °C to 30 °C/77 °F to 86 °F multiply the reading with the factor 0.6.
Absolute Humidity: 3 to 25 mg H2O / L

Reaction Principle
a) CH2CI2 + chromateä CI2
b) CI2 + amine ä red reaction product

Cross Sensitivity
100 ppm n-octane and 300 ppm carbon monoxide do not interfere the reading. In case of concentrations > 100 ppm n-octane, methylene chloride is not indicated. Other chlorinated hydrocarbons will be indicated.