Dräger-Tube Hydrogen Sulfide 0.2%/a - PN: CH28101
Dräger-Tube Hydrogen Sulfide 0.2%/a - PN: CH28101
$118.00
Dräger-Tube Hydrogen Sulfide 0.2%/a (10 Tests per Box)
PN: CH28101
| Application Range | |
| Standard Measuring Range | 0.2 to 7 vol.% |
| Number of Strokes n | 1 + 2 desorption strokes in clean air |
| Time for Measurement | approx. 2 min |
| Standard Deviation | ± 5 to 10% |
| Color Change | pale blue to black |
| Ambient Operating Conditions | |
| Temperature |
0 to 60 °C |
| Absolute Humidity | max 40 mg H2O/L |
| Cross Sensitivity |
| In the presence of sulfur dioxide, the indicating layer can change to a yellowish color, but the hydrogen sulfide measurement is not affected. Comparable concentrations of mercaptan will interfere with the reading. |
Dräger-Tube Hydrogen Sulfide 0.2%/a (10 Tests per Box)
PN: CH28101
| Application Range | |
| Standard Measuring Range | 0.2 to 7 vol.% |
| Number of Strokes n | 1 + 2 desorption strokes in clean air |
| Time for Measurement | approx. 2 min |
| Standard Deviation | ± 5 to 10% |
| Color Change | pale blue to black |
| Ambient Operating Conditions | |
| Temperature |
0 to 60 °C |
| Absolute Humidity | max 40 mg H2O/L |
| Cross Sensitivity |
| In the presence of sulfur dioxide, the indicating layer can change to a yellowish color, but the hydrogen sulfide measurement is not affected. Comparable concentrations of mercaptan will interfere with the reading. |