Dräger-Tube Hydrochloric Acid 0.2/a - PN: 8103481
Dräger-Tube Hydrochloric Acid 0.2/a (10 Tests per Box)
PN: 8103481
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of hydrochloric acid and has the following standard measuring ranges: 0.2-3 ppm 3-20 ppm Measurements: 10
| Application Range | |
| Standard Measuring Range | 0.2 to 3 ppm / 3 to 20 ppm |
| Number of Strokes n | 10 / 2 |
| Time for Measurement | approx. 2 min / 0.4 min |
| Standard Deviation | ± 10 to 15% |
| Color Change | blue to yellow |
| Ambient Operating Conditions | |
| Temperature |
5 to 40 °C |
| Absolute Humidity | < 15 mg H2O/L |
| Cross Sensitivity |
| No interference on the reading of 10 ppm H2S and 2 ppm SO2. Other acid gases are also indicated, however, with differing sensitivity. Chlorine changes the indicating layer to grey. If chlorine occurs simultaneously, HCl readings will be higher. |
Dräger-Tube Hydrochloric Acid 0.2/a (10 Tests per Box)
PN: 8103481
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of hydrochloric acid and has the following standard measuring ranges: 0.2-3 ppm 3-20 ppm Measurements: 10
| Application Range | |
| Standard Measuring Range | 0.2 to 3 ppm / 3 to 20 ppm |
| Number of Strokes n | 10 / 2 |
| Time for Measurement | approx. 2 min / 0.4 min |
| Standard Deviation | ± 10 to 15% |
| Color Change | blue to yellow |
| Ambient Operating Conditions | |
| Temperature |
5 to 40 °C |
| Absolute Humidity | < 15 mg H2O/L |
| Cross Sensitivity |
| No interference on the reading of 10 ppm H2S and 2 ppm SO2. Other acid gases are also indicated, however, with differing sensitivity. Chlorine changes the indicating layer to grey. If chlorine occurs simultaneously, HCl readings will be higher. |