Dräger Tube Chloroprene 5/a - PN: 6718901
Dräger Tube Chloroprene 5/a (10)
PN: 6718901
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of chloroprene and has the following standard measuring range: 5-60 ppm Measurements: 10
Application Range
Standard Measuring Range: 5 to 60 ppm
Number of Strokes n: 3 + 3 desorption strokes in clean air.
Time for Measurement: approx. 3 min
Standard Deviation: ± 10 to 15 %
Color Change: violet ä yellow brown
Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: 3 to 15 mg H2O / L
Reaction Principle
CH2=CH-CCI=CH2 + MnO4 – ä MnIV + various oxidation products
Cross Sensitivity
Many other organic compounds with C=C double bonds are indicated as well, but with different sensitivities. It is impossible to differentiate them. It is impossible to measure chloroprene in the presence of dialkyl sulfides.
Dräger Tube Chloroprene 5/a (10)
PN: 6718901
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of chloroprene and has the following standard measuring range: 5-60 ppm Measurements: 10
Application Range
Standard Measuring Range: 5 to 60 ppm
Number of Strokes n: 3 + 3 desorption strokes in clean air.
Time for Measurement: approx. 3 min
Standard Deviation: ± 10 to 15 %
Color Change: violet ä yellow brown
Ambient Operating Conditions
Temperature: 10 to 30 °C
Absolute Humidity: 3 to 15 mg H2O / L
Reaction Principle
CH2=CH-CCI=CH2 + MnO4 – ä MnIV + various oxidation products
Cross Sensitivity
Many other organic compounds with C=C double bonds are indicated as well, but with different sensitivities. It is impossible to differentiate them. It is impossible to measure chloroprene in the presence of dialkyl sulfides.