Dräger Tube Acetone 100/b - PN: CH22901
Dräger Tube Acetone 100/b (10)
PN: CH22901
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of acetone and has the following standard measuring range: 100-12000 ppm Measurements: 10
Application Range
Standard Measuring Range: 100 to 12,000 ppm
Number of Strokes n: 10
Time for Measurement: approx. 4 min
Standard Deviation: ± 15 to 20 %
Color Change: pale yellow ä yellow
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: 5 to 20 mg H2O / L
Reaction Principle
Acetone + 2.4-Dinitrophenylhydrazine ä yellow hydrazone
Cross Sensitivity
Other ketones are indicated but with different sensitivities. Aldehydes are indicated, but not esters. Ammonia causes the indicating layer to turn yellowish brown.
Dräger Tube Acetone 100/b (10)
PN: CH22901
Dräger-Tubes™ are used together with a Dräger tube pump for measurements of gases, vapors and aerosols, usually in air, and are one of the classic methods of gas analysis. This Dräger-Tube is used for the detection of acetone and has the following standard measuring range: 100-12000 ppm Measurements: 10
Application Range
Standard Measuring Range: 100 to 12,000 ppm
Number of Strokes n: 10
Time for Measurement: approx. 4 min
Standard Deviation: ± 15 to 20 %
Color Change: pale yellow ä yellow
Ambient Operating Conditions
Temperature: 0 to 40 °C
Absolute Humidity: 5 to 20 mg H2O / L
Reaction Principle
Acetone + 2.4-Dinitrophenylhydrazine ä yellow hydrazone
Cross Sensitivity
Other ketones are indicated but with different sensitivities. Aldehydes are indicated, but not esters. Ammonia causes the indicating layer to turn yellowish brown.