Dräger Tube Methanol 20/a - PN: 8103801
Rated 3.5/5 based on 11 customer reviews

Dräger Tube Methanol 20/a - PN: 8103801

$101.00
Image

Dräger Tube Methanol 20/a (10)
PN: 8103801

Application Range:

  • Standard Measuring Range: 20 to 250 ppm / 200 to 5000 ppm
  • Number of Strokes n: 15 / 5
  • Time for Measurement: approx. 6 min / approx. 2 min
  • Standard Deviation: ± 10 to 25 %
  • Color Change: Yellow → Mint Green

Ambient Operating Conditions:

  • Temperature: 15 to 30 °C
  • Absolute Humidity: ≤ 15 mg H2O / L

Reaction Principle:

  • Methanol + organometallic compound → green reaction product

Cross Sensitivity:

  • The tube does not differentiate between different alcohols.
  • Higher molecular alcohols are indicated with significantly decreasing sensitivity.
  • Ethers and xylene are indicated as well, but with different sensitivity.
  • ≤ 25 ppm formaldehyde, ≤ 50 ppm acetaldehyde, and ≤ 50 ppm toluene are not indicated.
  • Aliphatic petroleum hydrocarbons, ketones, esters, halogenated hydrocarbons and benzene are not indicated.

Dräger Tube Methanol 20/a (10)
PN: 8103801

Application Range:

  • Standard Measuring Range: 20 to 250 ppm / 200 to 5000 ppm
  • Number of Strokes n: 15 / 5
  • Time for Measurement: approx. 6 min / approx. 2 min
  • Standard Deviation: ± 10 to 25 %
  • Color Change: Yellow → Mint Green

Ambient Operating Conditions:

  • Temperature: 15 to 30 °C
  • Absolute Humidity: ≤ 15 mg H2O / L

Reaction Principle:

  • Methanol + organometallic compound → green reaction product

Cross Sensitivity:

  • The tube does not differentiate between different alcohols.
  • Higher molecular alcohols are indicated with significantly decreasing sensitivity.
  • Ethers and xylene are indicated as well, but with different sensitivity.
  • ≤ 25 ppm formaldehyde, ≤ 50 ppm acetaldehyde, and ≤ 50 ppm toluene are not indicated.
  • Aliphatic petroleum hydrocarbons, ketones, esters, halogenated hydrocarbons and benzene are not indicated.